Skip to main content

Table 1 Summary of results from previous BTNX rapid response fentanyl test strip (20 ng/mL cutoff) evaluations

From: Testing the test strips: laboratory performance of fentanyl test strips

Study

Year

Fentanyl limit of detection (ng/mL)

Fentanyl analog cross-reactivity

Interferences

Green et al.

2020

100

2 of 2 assessed cross-reacted

Not Assessed

Bergh et al.

2021

50

25 of 28 assessed cross-reacted

None found among substances assessed

Lockwood et al.

2021

25

Not Assessed

Methamphetamine, MDMA, Diphenhydramine

Wharton et al.

2021

100

19 of 29 assessed cross-reacted

Not Assessed

Park et al.

2022

200

13 out of 17 assessed cross-reacted

Not Assessed